Reference solution for testing your ORP electrode: 220 mV setpoint, 250 ml. So you can be sure that your redox readings are correct.

Out of stock

Manufacturer: APERA INSTRUMENTS (Europe) GmbH
Wilhelm-Muthmann-Straße 18
42329 Wuppertal
Germany
E-Mail: [email protected]

The ORP standard solution 220 mV is used for checking and regenerating ORP electrodes.
To check functionality, immerse the electrode in the ORP solution in a separate vessel. The measured mV value must not deviate by more than ± 15 mV from the value stated on the bottle (at your temperature). If the value is outside this ± 15 mV range, leave the electrode in the 220 mV standard solution for 15–30 minutes. This prepares the electrode surface for measurements and shortens the response time. (Product image may vary.)

  • Content: 250 ml standard solution with an ORP value of 220 mV (at 25 °C)

  • A table with ORP values at different temperatures can be found on the label.

  • Ready to use immediately: You don’t need to dissolve any powder in distilled water; you can use the solution directly.